Contact us

Suzhou Pullwin Auto Technology Co.,Ltd

Building 3, Meijite Industrial Product Expo Center No.1055, East Jinyang Road, Lujia Town Kunshan City

Wendy

0086-159-5093-1259

pullwin@pullwinsz.com

Fixed Sample Gas Condenser series

It is specially designed for gas analysis and pretreatment system, which not only can ensure effective removal of water from sample gas, but also ensure the minimum loss of components in the sample gas。

Features

☆ High quality TEC refrigeration sheet, better gas processing capacity

☆ Built-in humidity alarm module, which connect with pump,  can avoid analyzer damage effectivly

☆ PVDF needle valve anticorrosive flowmeter, free to adjust the suction flow rate

☆ Lightweight design, more suitable for outdoor site

Model No.PW-GC-F series
Working temprature5~50℃;
Inlet dew point≤80℃(working temprature 25℃)
Outlet dew point≤4℃(working temprature 25℃)
Refrigerating control temperature3℃ ( factory settings )
Cooling modefluorine free compressor
Condenser pipeglass heat exchange tube
Input sample gas pressure≤3bar
Differential air pressure≤3mbar(flow rate 2Nl/min.)
Sample gas processing capacity250NL/h
Sample gas inlet/outletOD6mm 
Water-drop outletOD6mm
Warm-up time≤30min
Alarmhigh/low termprature alarm    NC/NO alarm

Fixed-sample-gas-conderser.jpg