This system provides users with flow measurement solutions beyond traditional methods such as matrix pilot tubes and thermal balance, especially in sites of low vel- ocity, wet flue gas and corrosive exhaust gas ( waste power plant).
Feature
☆ Can achieve high precision measurement at very low flue gas velocity
☆ On-line automatic calibration function of the whole system
☆ Wide measuring range to meet different field condition
☆ With purging isolation unit,low maintenance
☆ Easy installation & maintenance, can install two-side or single side
☆ Titanium probe can be used for corrosive smoke
Principle
This system measures the flue gas emission according to the principle of ultraso-
nic propagation time difference. The ultrasonic sensor transmits waves in opposite dir-
ections to each other, then ultrasonic waves in the two directions produce a time diffe-
rence. The flow rate is calculated according to the difference.
The flow velocity measurement is related to the up-flow transmission time T1, th-
e backward flow transmission time T2,the length of the transmission path and the ang-
le between the smoke flow. It has nothing to do with temperature, pressure,compositi-
on of the flue gas.
According to the transmission speed of ultrasonic reflue gas, the temperature of
flue gas is calculated. Combined with the cross-sectional area of flue, the volume flow
rate of flue gas is calculated.
System composition
This system consists of two transmitters ( receivers) and control units.The control
unit is used for signal input and output, for calculating volume flow to reference condi-
tions and for rapid processing via LC display.
Installation of transmitter ( receiver ) unit
1. Two-side installation: two transmitter (receiver) units are installed on either side
of the flue at a specific angle to the direction of the gas flow.
2. Single side installation: only one transmitter(receiver) is installed on the flue at
a specific angle to the direction of the gas flow.
Measuring object Flue gas velocity,Volume flow,Temperature,Pressure
Measuring range 0~40m/s
Flue diameter 0.14~13m
Resolution ratio 0.001m/s
Detection limit ≤0.1m/s
RSD ±0.1m/s
Zero drift ≤±1%
Response time 5s(adjustable)
Working conditions
Flow gas temperature -40~450℃
Pressure -6.1~10kPA
Humidity 0~100%RH
Environment temperature -40~60℃
Electronic Control Unit
Output 4~20mA 4 ways
Communication RS232
Power supply 220V 50HZ
Grade IP65 Ex dbIIB T4Gb